Description
Automatic microplate reader is a professional instrument for reading and analyzing the results of enzyme-linked immunoassay (EIA) experiments.This product is widely used in applied research in biology, agricultural science, food and environmental science.
This product has the following features:
- Adopt 7-inch high-resolution LCD capacitive touch screen, no keyboard required, simple and intuitive operation, easy to use.
- 8-bit filter wheel, with 4 filters as standard; Other filters are optional; The wavelength range is 340-750nm, which can meet various applications;
- 9-channel vertical light path; Automatic positioning function of the center of the enzyme labeling hole.
- It has the function of single wavelength and dual wavelength detection, and the plate reading is fast; it only takes 10 seconds to read the entire 96-well plate or 16 seconds to complete the dual wavelength detection.
- The built-in software can realize the detection of kinetics, standard curve, qualitative and quality control; The software includes program generation and storage, as well as detection data storage functions, with large storage capacity; it also has powerful curve fitting, kinetic analysis and reporting functions.
- It has the functions of self-checking and diagnosing the optical path, moving position, etc.
- It has the function of vibrating plate, and the vibrating plate time and speed are adjustable.
- The energy-saving design of the light source maximizes the service life of the light source.
- A variety of data output interfaces are available to facilitate data transmission and processing.
- Measurement data can be directly exported to Excel for further analysis.
Specifications
Model | MPR-A9600/MPR-A9600T |
Light source | 6V 10W halogen lamp |
Wavelength range | 340~750nm |
Filter | Standard configuration: 405, 450, 492, 630nm four filters;It can be loaded with up to eight filters |
Absorbance range | 0.000-4.000Abs |
Resolution | 0.001Abs |
Linearity | (0, 2.0)≤±1%;[2.0, 4.0)≤±2% |
Wavelength indication error | ≤±3nm |
Wavelength repeatability | ≤±1.5nm |
Repeatability | [0, 3.0)≤0.3%, [3, 4.0)≤1% |
Stability | [0, 3.0)≤0.3%,[3.0, 4.0)≤1% |
Accuracy | [0, 2.0)≤ ± 0.005, [2.0 ,3.0)≤ ± 1%, [3.0, 4.0)≤ ± 1.5% |
Sensitivity | Sensitivity≥0.010 |
Channel difference | <0.01A(1.0Abs standard filter ) |
Measuring speed | Single wavelength <10s/96 wells, dual wavelengths <16s/96 wells (normal mode) |
Incubation temperature | Room temperature+4℃-50℃(MPR-A9600T) |
External dimensions(W×D×H) | 280x432x220 mm(W×D×H) |
Net weight (kg) | 10 |
Ambient temperature | 4℃~ 45℃ |
Relative humidity | ≤70% |
Power supply | VAC 100-240V |