High-Performance Liquid Chromatograph Chromaster
HPLC for today and tomorrow 1. Outstanding Performance Two performance capabilities supporting data reliability: excellent reproducibility made possible by the pump and autosampler and excellent stability of the column oven and detector. 2. Easy-to-use Along with simple operation and ease of use, Chromaster offers ease of maintenance, a critical requirement for HPLC. 3.Robustness Hitachi's reputation for instrument robustness and reliability continues with the Chromaster, which is built using stronger materials and manufactured with Hitachi's strict quality control standards.

Features

-

-

*1:Optional

*2: Autosamplers with or without thermostat are available

*3: Pumps with or without Auto-purge valve are available

-

Module lineup

Pump

5110: Pressure up to 40 MPa
5160: Pressure up to 60 MPa

Item51105160
Operating flow late range0.001 to 9.999 mL/min0.001 to 5.000 mL/min
Maximum operating pressure40 MPa (0.001 to 5.000 mL/min)
20 MPa (5.001 to 9.999 mL/min)
60 MPa (0.001 to 2.500 mL/min)
30 MPa (2.501 to 5.000 mL/min)
Dimensions340(W) × 440(D) × 140(H) mm
Number of mixed solvents*1Up to 4

*With low pressure gradient unit (optional)

Autosampler

5260: Withstand pressure 60 MPa, Loop injection method
5280: Withstand pressure 60 MPa, Direct injection method

Item52605280
Sample injection systemLoop injection method (Cut injection, All volume injection, Full loop injection method)Direct injection method
Withstand pressure60 MPa60 MPa
Washing functionEquipped with built-in washing pump
Two-solvent washing function
Wash solvent delivery by a syringe
Sample rack temperature control1 to 45 °C1 to 35 °C
(Temperature setting range)Capable of cooling and heatingCooling only
Sample capacityStandard120 × 1.5 mL200 × 1.5 mL
Optional72 × 4 mL128 × 4 mL

Column Oven

5310: Horizontal type
6310: Vertical type

Item53106310
Temperature setting range1 to 85 °C (1 °C step)4 to 90 °C (1 °C step)
Temperature control range[Ambient temperature -15 °C] to [Ambient temperature +60 °C] and range of the temperature setting[Ambient temperature -15 °C] to [Ambient temperature +75 °C] and within temperature setting range
Column capacity300 mm × 3 (Maximum)
Built-in 3-liter water tank-Available
Dimensions410(W) × 440(D) × 140(H) mm165(W) × 515(D) × 689(H) mm (Legs are not included)

Detector

5410: UV Detector
5420: UV-Vis Detector
5430: Diode Array Detector
5440: Fluorescence Detector
5450: RI Detector
5610: MS Detector

Item5410 UV Detector5420 UV-Vis Detector5430 Diode Array Detector
Optical systemDouble-beam ratio photometric systemSingle-beam ratio photometric system
Wavelength range190 to 600 nm190 to 900 nm190 to 900 nm
Flow cellOptical path length 10 mm
Simultaneous measurementTwo-wavelength*spectrum acquisition
Dimensions340(W) × 440(D) × 140(H) mm

Item5440 Fluorescence Detector
Light sourceXe lamp, Hg lamp for checking wavelength
Wavelength rangeEx: 200 to 850 nm
Em: 250 to 900 nm (Change photomultiplier at 731 nm or more)
Flow cellIrradiation volume 12 µL
Dimensions340 (W) × 440 (D) × 280 (H) mm
Item5450 RI Detector
Refractive index range1 to 1.75
Temperature control rangeOFF, and 30 to 50 °C
Dimensions340 (W) × 440 (D) × 140 (H) mm, excluding projections
Item5610 MS Detector
Measurement mass range (m/z)20 to 1,000
Ion sourceElectrospray ionization (ESI)
Dimensions440 (W) × 610(D) × 430 (H) mm
N2 gas usageMax flow rate 3.4 L/min, Pressure 300 ± 20 kPa

Organizer

ItemOrganizer
Output powerDC24 V, 450 W
Supplies power to one pump, one autosampler, one detector (UV detector, UV-Vis detector, Diode array detector, or RI detector), and one interface control board
Bottle capacity and space1.0 L bottle × 6 and 500 mL bottle × 3(Maximum), 314(W) × 280.8(D) mm
Dimensions340 (W) × 420 (D) × 200 (H) mm

Data

System Performance

Analysis Example from the Chromaster 60 MPa System

The 60 MPa system, which can be used with columns containing 2.0 µm or smaller particles, and core-shell columns, is capable of conventional HPLC analysis and also ensures improved resolution performance and shorter analysis time.

-
SampleAlkylphenones
ColumnLaChrom II C18 (5 µm) 4.6 mm I.D. × 150mmL
LaChromUltra II C18 (1.9 µm) 2.0 mm I.D. × 50 mmL
Column temperature40 °C
Mobile phaseA: H2O, B: CH3CN
Gradient modeHigh Frequency Mode
Wavelength247 nm

The following options are used with UHPLC column:

  • Low dispersion piping kit
  • Semi-micro mixer (200 µL)
  • Semi-micro flow cell unit

Example of Carry-over analysis (5260)

Even when highly adsorbing Chlorhexidine is analyzed, the carry-over is extremely low.

-
-

Easy to use

Analysis Example from the Chromaster 60 MPa System

Most optional accessories are internally mounted to reduce the HPLC system's height. The handle located on the front side of the organizer moves vertically for easy access to solvent bottles.

-

Module operations and the replacement of consumable and maintenance parts can be performed from the front side.

-

With attention to detail on the housing of tubes and wires, the system keeps tubes from getting tangled up, ensures the ease of replacement, and provides adequate seismic stability.

-
-

Column management system (optional)

Hitachi's column management system can manage the Log information on analytical columns and guard columns from any manufacturer.
Log information can be written and read through a connector mounted on the column oven or USB port in the computer. ID Tags can be used repeatedly*.

*Approximate read/write life time : 100,000 times

-
-

Wakeup (automatic pre-analysis tasks) and Sleep (automatic post-analysis tasks) programs

Automatic system Wakeup and Sleep from GUI

  • In Conditioning, up to 10 programs can be created by combining any of the module settings, such as pre-analysis tasks of system (Wakeup), and post-analysis tasks of system (Sleep).
  • For Wakeup program ending time, you can specify any time on current day, the following day, or two days later.
  • The Sleep program starts at a specified time on the current day/the following day, or after the end of a continuous analysis run.

Automated system standby can reduce the amount of time required to make preparations for an analysis run.

-

Applications

Here are two examples of experiments that can be done using the Chromaster system:

  1. Column switching valves enable automation of the pretreatment process such as biological sample enrichment, removing impurities and other processes.
  2. The ability to switch between two different wavelengths in real time can be used to resolve the co-elution of impurities in drug products.

PharmaceuticalFoodEnvironment

Pharmaceutical

Application for Biomedicines - Peptide Mapping Method - 

Analysis of Two New Drugs in the Japanese Pharmacopoeia Sixteenth Edition 

Analysis Example for a Drug Listed in the Japanese Pharmacopoeia Sixteenth Edition (Isoleucine, Leucine, and Valine Granules) 

Analysis Example for Drugs in the Japanese Pharmacopoeia Sixteenth Edition (Acetylcysteine) 

Automation of Pretreatment by Column-Switching 

Analysis of decomposition products in aspirin by application of two-wavelength detection 

Analysis of steroid drugs 

Analysis of anti-mycoticagents 

Simultaneous analysis of chlorzoxazoneand its metabolites 

Simultaneous Analysis of Testosterone and Metabolites 

Simultaneous Analysis of Phenoxyethanol and Parabens 

Simultaneous Analysis of Main and Trace Components by High-sensitivity DAD 

Analysis of Combination Drugs (Identification and Quantitative Analysis by DAD) 

Food

Analysis of Catechins 

Analysis of Resveratrol 

Analysis of Bis(2-ethylhexyl) phthalate(DEHP) in Drinking Water 

Analysis Example of Aflatoxin M1 

Highly Sensitive Analysis of Organic Acids 

Simultaneous Analysis of Fat-soluble Vitamins 

Simultaneous Analysis of Water-Soluble Vitamins 

Environment

Simultaneous analysis of four agricultural chemical components 

Analysis of anion surfactants 

Analysis of 2,4-DNPH-Derivatized Aldehydes 

Simultaneous Analysis of Main and Trace Components by High-sensitivity DAD 

NOTE:
These data are an example of measurement; the individual values cannot be guaranteed.
The system is for research use only, and is not intended for any animal or human therapeutic or diagnostic use.