Graphite Furnace System, GRD-3202
Heating steps: 9 steps
Temperature range (nominal Temperature): 20℃~3000℃
Slope heating time: 0~999s
Heating holding time: 1s~999s (the sum of both time should be less than 999 seconds)
Inert gas needed: argon, pressure of entry larger than 0.3Mpa
Cooling water: tap water or cycling water, flow rate no less than 2L/min

Description

GRD-3202 graphite furnace system is a system specially used for the quantitative analysis of trace elements in substances by graphite furnace atomic absorption method. It is an important optional accessory for SP-IAA320 atomic absorption spectrophotometer.

Specifications

ModelGRD-3202
Heating steps9 steps
Temperature range (nominal Temperature)20℃~3000℃
Slope heating time0~999s
Heating holding time1s~999s (the sum of both time should be less than 999 seconds)
Inert gas neededArgon, pressure of entry larger than 0.3Mpa
Cooling waterTap water or cycling water, flow rate no less than 2L/min
OutputLCD
Power source220V±22V, 50Hz±1Hz, 3A 220V±22V, 50Hz±1Hz, 30A
Power5.1kW for 220V on about 2700℃
Ambient temperature+10℃~+30℃
Relative humidityLess than 85%
Instrument rating power7.2kW
Gas pressure alarm, furnace overheated alarm
Interface with Atomic absorption spectrophotometer, autosampler and RS232
High-power temperature heating function (1000℃~2700℃)
Test sample together with atom absorption equipment. Cd ≤1×10^-12g; Cu ≤1×10^-10g