DALPHIL DUNAMIS LTD
Visit our medical equipment website 
Here
Enquiries
Class II Type B2 Biosafety Cabinets, BSC-IIB2-3J
Cleanliness level: ISO level 4
Classification: Class II Type B2
Air flow mode: 100% discharge, 0 internal circulation
Vertical laminar negative pressure model
Working zone size: 1000*520*640 mm

Description

BSC series biological safety cabinets (biosafety cabinet) are biological safety isolation equipment used in biological safety laboratories or other laboratories to protect personnel, tested samples and the environment. It can meet the operation of pathogenic body with risk level of 1, 2 and 3.

High-Quality&Ergonomic Construction

  • The outer box is made of high-quality cold-rolled steel plate with electrostatic spraying;
  • The three sides of the operation area are made of stainless steel integrated structure;
  • Internal washable parts are treated with 10mm large rounded corners, which is easy to clean;
  • 10° inclined design, making the operator more comfortable;
  • The exhaust air is filtered by a high-efficiency filter to protect the environment from pollution;

Energy-efficient

  • Class II Type B2 Biosafety Cabinets;
  • Vertical laminar negative pressure model, Air flow mode: 100% discharge, 0 internal circulation;

High security

  • Negative pressure design ensures the safety of operators;
  • Integrated sliding safety door positioning, easy to operate;
  • Equipped with filter blocking alarm, blower overload alarm, working window opening limit alarm system;

Specifications

ProductBiological safety cabinet (biosafety cabinet)
ModelBSC-IIB2-3J
Cleanliness levelISO level 4
Filtration efficiency≥99.9995%,@0.12μm
ClassificationClass II Type B2
Average flow rate of downdraft(m/s)0.25~0.40
Average flow rate of inflow air(m/s)≥0.5
Sound emission (dBA)≤67
Vibration peak (μm)≤5
ElectricityAC single-phase 220V/50Hz
Air balance biological protectionPersonnel protection:
a.The total number of colonies in the impact sampler:≤10CFU/time
b.Total colony count of the slot sampler:≤5CFU/time
Product protection 1-8CFU/ml(repeat 3 times, 5min/time) total number of colonies:≤5CFU/time
Cross contamination protection 1-8CFU/ml(repeat 3 times, 5min/time) total number of colonies:≤5CFU/time
Max power(KV·A)≤1.5 (Spare socket included)
Weight(Kg)230/240
Working zone size( W*D*H)1000*520*640 mm
External dimensions(W*D*H)1165*760*2200 mm
Fluorescent lamp20W*2pcs
UV lamp18W*1pc
Illumination (Lux)900
Supply and exhaust pipe diameter(mm)Φ250
Wind directionTop in,top out
Scroll to Top