DALPHIL DUNAMIS LTD
Visit our medical equipment website 
Here
Enquiries
Class II Type A2 Biosafety Cabinets, BSC-IIA2-2J
Cleanliness level: ISO level 4, Class II Type A2
Working zone area: 600*500*520 mm
Unique negative pressure leak-proof design
Equipped with height limit alarm device
Front window and germicidal lamp interlocking design

Description

Biological safety cabinet is necessary equipment in the laboratory in the search of microbiology, biomedical, DNA recombinant, animal experiment and biological products, especially in the occasion that operator need to adopt protective measure, such us medical and health, pharmacy, medical research.

Features

High-Quality&Ergonomic Construction

  • The outer box is made of high-quality cold-rolled steel plate with electrostatic spraying;
  • The three sides of the operation area are made of stainless steel integrated structure;
  • Internal washable parts are treated with 10mm large rounded corners, which is easy to clean;
  • 10° inclined design, making the operator more comfortable;

Energy-efficient

  • Class II biological safety cabinet.
  • Airflow system: 30% air exhaust,70% air recirculation.
  • Two high-efficiency filters are standard;

High security

  • Equipped with height limit alarm device.
  • Unique negative pressure leak-proof design;
  • The LCD blue screen digital indicates the value of each parameter, and the password is used to modify the program setting to prevent misoperation.

Convenient and practical front window design

  • With integrated sliding door that can be positioned anywhere and can be fully closed for sterilization
  • Adopts safety tempered glass not less than 6mm thick.
  • Front window and germicidal lamp interlocking design: when the front window is opened, the germicidal lamp goes out.

Specifications

ModelBSC-IIA2-2J
ClassificationClass II Type A2
Cleanliness level ISO level 4
Filtration efficiency≥99.9995%,@0.12μm
Downflow air velocity0.33m/s±0.025
Inflow air velocity0.53m/s±0.025
Sound emission≤67dBA
Vibration semi-peak value≤5μm
Electricity/Max. powerAC, 220V/50Hz / 1kW
Biological safety personnel protectionSampling number of all impact samplers CFU ≤ 10, sampling number of slit samplers CFU≤5
Product protectionSampling number of all Petri dishes CFU≤5
Cross contamination protectionSampling number of all Petri dishes CFU≤2
Working zone size600*500*520 mm
External dimension740*650*1850 mm
Illumination≥900Lux
Fluorescent lamp20W*2pcs
UV lamp18W*1pc
Weight140kg
Applicable number of people1
Wind directionTop out