DALPHIL DUNAMIS LTD
Visit our medical equipment website 
Here
Enquiries
Benchtop Multi-parameter Analyzer
Color High contrast LCD touchscreen.
Data Storage1000 sets for each parameter.
IP54 waterproof.
Support for USB or RS-232 communication.

Description

The BEP-M500T Benchtop Multi-parameter Analyzer a multi-purpose measurement method, which is suitable for the fields of health disease control, environmental protection, biological agriculture and forestry, water quality analysis, scientific research and institutions of higher learning.

General Features

  1. Color High contrast LCD touchscreen, 7.0 inches.
  2. Intelligent operation system provides features including user management, calibration management, method management, data management and etc.
  3. Multi-reading feature allows auto-read, timed-read and continuous-read.
  4. Automatic/Manual temperature compensation ensures accurate results.
  5. Data Storage1000 sets for each parameter (GLP-compliant).
  6. Data analysis feature helps user review, compare and recalculate results.
  7. Support for USB or RS-232 communication.
  8. Reset feature automatically resumes all settings back to factory default options.
  9. IP54 waterproof.

pH

  1. 1-6 points calibration with Standard Recognition.
  2. Selectable pH buffer groups, including DIN, NIST, USA, MERK, JIS, GB.
  3. Support for pH standard group customization.
  4. Automatic electrode diagnosis with pH slope and offset display.

Ion

  1. 1-6 points calibration
  2. Selectable measurement unit, including μg/L, mg/L, g/L, mmol/L, pX, ppm, etc.
  3. Multi-measurement modes are supported, including Direct Reading mode, Standard Addition mode, Sample Addition mode and GRAN mode.
  4. Over 10 methods are built-in, including F, Cl, Br, I, NO3, BF4, NH4+, K+, Na+, Ca2+, Cu2+, Pb2+, Ag+and etc., user-defined method is supported.

Conductivity

  1. 1-5 points calibration with Standard Recognition.
  2. Settable parameters, including cell constant, reference temperature (5/10/15/18/20/25°C), temperature compensation coefficient and TDS factor.
  3. Temperature compensation type (none, linear, pure water).

DO

  1. Support for air-saturated water or zero oxygen calibration.
  2. Auto barometric pressure compensation.
  3. Manual Salinity Factor Correction.
  4. Selectable pressure unit, including kPa, mbar, Torr, Atm.

Specifications

ModelBEP-M500T
ParameterspH/EC/ISE/DO/Temp. (mV/ORP/pX/Resistivity/TDS/Sal./DO Saturation)
pHRange-2.000 to 20.000 pH
Resolution0.1, 0.01, 0.001pH
Accuracy±0.002 pH
Calibration PointsUp to 6
Standard CustomizationYes
Calibration ReminderYes
Standard RecognitionNIST, GB, DIN, USA, MERK, JIS buffers
Slope LimitYes
mV/ORPRange-2000.00 to 2000.00 mV
Resolution0.1, 0.01 mV
Accuracy±0.1 mV or ±0.03% ofreading whichever is greater
EH ORP ModeYes
ORP Calibration Points1 custom point (relative mV)
pXRange-2.000 to 20.000 pX
Resolution0.1, 0.01, 0.001 pX
Accuracy±0.002 pX
Calibration PointsUp to 6
ISERange1E-9 to 9.999E9
Unitsmol/L, mmol/L, g/L, mg/L, μg/L, ppm
ResolutionUp to 4 significant digits
Accuracy±0.3%
Calibration PointsUp to 6
ConductivityRange0.000 μS/cm to 2000 mS/cm
Resolution0.001 μS/cm minimum,various with range selection
Accuracy±0.5% FS
Reference Temperature5, 10, 15, 18, 20, 25 ºC
Calibration PointsUp to 5
Calibration ReminderYes
Standard Recognition10 μS/cm, 84 μS/cm, 500 μS/cm, 1413 μS/cm, 12.88 mS/cm
ResistivityRange5.00 Ω·cm~100.00 MΩ·cm
Resolution0.01 Ω·cm minimum
Accuracy±0.5% FS
TDSRange0.000~1000 ppt
Resolution0.001 ppm minimum, changed with range
Accuracy±0.5%FS
SalinityTypeNaCl%
Range(0.00~8.00)%
Resolution0.01%
Accuracy±0.1%
DissolvedOxygen Concentration – PolarographicRange0.00 to 99.99 mg/L
Resolution0.01 mg/L
Accuracy±0.3mg/L (0.00 to 20.00 mg/L);±10% (20.00 to 99.99 mg/L)
Calibration ReminderYes
Calibration PointsAir-saturated water or zero point
Barometric CompensationYes
Manual Salinity Factor CorrectionYes
% Saturation -PolarographicRange (0.0 to 600)%
Resolution0.10%
Accuracy±2.0%
TemperatureRange-10 to 135 ºC, 14 to 275 ºF
UnitºC, ºF
Resolution0.1
Accuracy±0.1
MeasurementReading ModeAuto Read (Fast, Medium, Slow), Timed, Continuous
Reading PromptsReading, Stable, Locked
Temp. CompensationATC, MTC
Data managementData Storage1000 Groups
GLP FeaturesYes
Log ManagementYes
InputspH ElectrodeBNC(Q9)
DO with Temp. Probe4-pin aviation connector
Conductivity with Temp. Probe5-pin aviation connector
OutputsUSBUSB 2.0 flash memory device, PC
RS 232printer, scanner, auto-sampler
Display optionsBacklightYes
Auto Shutdown1~60 min, off
IP RatingIP54
Date and TimeYes
GeneralElectricityAC Adapter, 100-240V AC input, DC24V output
Dimensions220×195×68 mm
Weight950g