Description
- Horizontal circular inner chamber and jacket structure, safety, heat preservation, and drying function;
- Disc type embedded gear and Electric lock door structure,safety and reliable;
- Door seals are pneumatic or self-expanding
- Equipped with standard equipment verification interface
- Pipeline system for the optimal design, SUS 304 pipeline, clamp structure, quick assembly
- Control system: HMI man-machine interface, touch screen + microcomputer;
- FO value and temperature time double guarantee sterilization effect, with perfect sterilization records;
- With program selection, parameter setting, equipment operation, report processing and traceability Functions;
- Multiple safety protection(over-temperature, over-pressure protect)
- Sterilization process: fabric, equipment, liquid, BD, pressure testing, custom defined multiple procedures
- Built-in printer
- USB function
Specifications
Model | Volume(L) | Overall Dimensions (L×W×H)mm | Chamber Size (Ø×L)mm | Steam Consumption (KG) | Water Consumption (KG) | Power (KW) | Net Weigh (KG) |
STH-N100D | 100 | 1160×650×1680 | Ø420×750 | ≤10 | ≤30 | ≤2+⑨ | 250 |
STH-N120D | 120 | 1290×650×1680 | Ø420×880 | ≤11 | ≤30 | ≤2+⑨ | 280 |
STH-N150D | 150 | 1290×690×1680 | Ø472×880 | ≤12 | ≤190 | ≤2+⑫ | 350 |
STH-N200D | 200 | 1560×690×1680 | Ø472×1150 | ≤13 | ≤190 | ≤2+⑫ | 370 |
STH-N300D | 300 | 1430×850×1760 | Ø620×995 | ≤18 | ≤200 | ≤2+⑮ | 450 |
STH-N400D | 400 | 1755×850×1760 | Ø620×1320 | ≤22 | ≤200 | ≤2+⑮ | 510 |
Main Technical Parameter
Design pressure | -0.1~0.3Mpa |
Design temperature | 144℃ |
Working pressure | 0.245Mpa |
Working Temperature | 105-138℃ |
Vacuum | -0.086Mpa |
Temp accuracy | ±1℃ |
Temperature Equilibrium | <+2.0℃ |
Water Pressure | 0.1-0.2Mpa |
Compressed air pressure | 0.4~0.7Mpa |
Steam Pressure | 0.3-0.7Mpa |
Power voltage | Three-phase five-wire AC 380V 50HZ |